Features:
Expanded Ranges - HI98197 has an expanded conductivity range from 0.000 S/cm to 1000 mS/cm actual conductivity (400 mS/cm temperature compensated). The HI98197 also has ranges for TDS, resistivity and three salinity scales. This meter automatically recognizes the probe type (two or four ring) and allows the user to adjust the nominal cell constant.
USP <645> - HI98197 can be used to perform all three stages of USP <645> method required for EC measurement of ultrapure water and generates a report when the any of the three stages are met.
5-Point Calibration - Calibration can be performed up to five points with seven preprogrammed conductivity standards. For salinity (% NaCl range), HI7037 standard allows users to perform a one-point calibration.
Temperature Compensation - EC and TDS measurements are fully customizable and include: cell constant selection between 0.0 and 10.000 and a selection of linear, non-linear (natural water) or no temperature compensation. The temperature compensation coefficient range is configurable from 0.00 to 10.00%/C with a choice of reference temperatures of 15C, 20C and 25C.
TDS Conversion Factor - The conductivity to TDS factor is adjustable from 0.40 and 1.00.
User Profiles - The HI98197 allows for the creation of 10 measurement profiles with various user settings including temperature compensation, range selection, cell constant, and TDS conversion factors.
Logging - The log-on-demand mode allows the user to record and save up to 400 readings while the lot logging at selected intervals can store up to 1000 readings. This data can then be transferred to a PC with Hannas HI920015 micro USB cable and HI92000 software.
Good Laboratory Practice (GLP) - GLP data includes date, time, standards used, temperature compensation used, reference temperature, and TDS conversion factor (TDS) is directly accessible by pressing the dedicated GLP key.
Contextual Help - Access Help at any time at the press of a dedicated button and view content specific information based on the screen that is currently being viewed.
Graphic LCD - The high contrast graphic LCD is easy to read outdoors in bright sunlight as well as in low-lit areas with the backlight. A combination of dedicated and virtual keys allows easy, intuitive operation in a choice of languages.
Thermoformed Carrying Case - The compact, durable HI720197 carry case is thermoformed to hold all necessary components for taking a field measurement, including the meter and probe, beakers and conductivity standards.
Specifications
EC Range: 0.000 to 9.999 S/cm; 10.00 to 99.99 S/cm; 100.0 to 999.9 S/cm; 1.000 to 9.999 mS/cm; 10.00 to 99.99 mS/cm; 100.0 to 1000.0 mS/cm (actual conductivity; temperature compensated to 400 mS/cm)
EC Resolution: 0.001 S/cm; 0.01 S/cm; 0.1 S/cm; 0.001 mS/cm; 0.01 mS/cm; 0.1 mS/cm
EC Accuracy: 1% of reading (0.01 S/cm or 1 digit, whichever is greater)
EC Calibration: Automatic up to five points with seven memorized standards (0.00 S/cm, 84.0 S/cm, 1.413 mS/cm, 5.00 mS/cm, 12.88 mS/cm, 80.0 mS/cm, 111.8 mS/cm)
TDS Range: 0.00 to 99.99 ppm; 100.0 to 999.9 ppm; 1.000 to 9.999 ppt (g/L); 10.00 to 99.99 ppt (g/L); 100.0 to 400.0 ppt (g/L)
TDS Resolution: 0.01 ppm; 0.1 ppm; 0.001 ppt (g/L); 0.01 ppt (g/L); 0.1 ppt (g/L)
TDS Accuracy: 1% of reading (0.05 mg/L (ppm) or 1 digit, whichever is greater)
TDS Calibration: Based on conductivity calibration
Resistivity Range: 1.0 to 99.9 cm; 100 to 999 cm; 1.00 to 9.99 Kcm; 10.0 to 99.9 Kcm; 100 to 999 Kcm; 1.00 to 9.99 Mcm; 10.0 to 100.0 Mcm
Resistivity Resolution: 0.1 cm; 1 cm; 0.01 Kcm; 0.1 Kcm; 1 Kcm; 0.01 Mcm; 0.1 Mcm
Resistivity Accuracy: 1% of reading (10 or 1 digit, whichever is greater)
Resistivity Calibration: Based on conductivity calibration
Salinity Range: % NaCl: 0.0 to 400.0%; practical salinity: 0.00 to 42.00 (PSU); natural seawater scale - UNESCO 1966: 0.00 to 80.00 (ppt)
Salinity Resolution: 0.1%; 0.01
Salinity Accuracy: 1% of reading
Salinity Calibration: Max. one point only in % range (with HI7037 standard); use conductivity calibration for all other ranges
Temperature Range: -20.0 to 120.0C (-4.0 to 248.0F)
Temperature Resolution: 0.1C (0.1F)
Temperature Accuracy: 0.2C; 0.4F (excluding probe error)
Temperature Calibration: One or two points
Temperature Compensation: noTC, Linear (-20 to 120 C; -4 to 248F), Non Linear - ISO/DIN 7888 (-0 to 36 C; 32 to 96.8F)
Measurement Modes: Autorange, autoend, lock and fixed range
Reference Temperature: 15C, 20C and 25C
Temperature Coefficient: 0.00 to 10.00 %/C
TDS Factor: 0.40 to 1.00
Electrode/Probe: HI763123 platinum, four ring conductivity/TDS probe with internal temperature sensor and 1 m cable (included)
Log-on-demand: 400 samples; lot logging: 5, 10, 30 sec, 1, 2, 5, 10, 15, 30, 60, 120, 180 min (max 1000 samples)
Memorized Profiles: Up to 10
PC Connectivity: Opto-isolated USB with optional HI 92000 software and micro USB cable
Battery Type/Life: 1.5V AA batteries (4) / approximately 100 hours of continuous use (without backlight), 25 hours with backlight
Auto-off: User selectable: 5, 10, 30, 60 min or can be disabled
Environment: 0 to 50C (32 to 122F); RH 100% (IP67)
Dimensions: 185 x 93 x 35.2 mm (7.3 x 3.6 x 1.4)